Last reviewed: November 27, 2026

Superagent · Complizen's FDA AI

Ask about your device. Get a cited answer in seconds.

No. A TİTCK registration, its ÜTS record and the CE mark behind it have no legal standing at FDA, and no US and Turkey agreement changes that. What helps with FDA clearance is the evidence in your technical file: test reports to FDA-recognized standards, your risk file and your ISO 13485 quality system.

Key takeaways

  • FDA's mutual recognition agreements are with the EU, Switzerland and the UK, and they cover drug inspections. None covers medical devices, and none is with Turkey [9].
  • Turkey's Tıbbi Cihaz Yönetmeliği (Medical Device Regulation), Official Gazette No. 31499 of June 2, 2021, follows EU MDR 2017/745 and requires CE marking. It does not mention FDA [1].
  • Of the 11 technical file items in our table, 7 can carry over to a 510(k) in full or in part, such as test reports and the risk file. The other 4 cannot: the Clinical Evaluation Report, the notified body certificate, the ÜTS record and the EU equivalent device.
  • FDA cleared 28 510(k)s for Turkey-based applicants from January 1, 2023 to November 26, 2026, all through the Traditional 510(k) route [11].
  • FDA registration, device listing and a US Agent are needed before a device is imported, not before 510(k) clearance [4].

Does a TİTCK registration help with FDA clearance?

Not as a credential. FDA makes its own decision on your device, and a TİTCK registration carries no weight in that decision.

When Turkish teams say "TİTCK registration", they usually mean two things. The first is the CE marking that lets the device be sold in Turkey. The second is the device's record in Ürün Takip Sistemi (ÜTS), Turkey's product tracking system [2]. ÜTS is run by Türkiye İlaç ve Tıbbi Cihaz Kurumu (TİTCK), Turkey's medicines and medical devices agency. Neither is a review that FDA relies on.

FDA treats even its own registration this way. Under 21 CFR 807.39, FDA registration "does not in any way denote approval of the establishment or its products" [4]. A foreign registration counts for even less. What FDA reviews is your evidence, and much of that evidence already sits in your technical file. Our guide to what transfers from a CE mark to a 510(k) covers the EU side in detail. This page covers what is specific to Turkey.

How does Turkey regulate medical devices?

TİTCK regulates medical devices under the Tıbbi Cihaz Yönetmeliği, published in the Resmî Gazete (Official Gazette) No. 31499 (Mükerrer) on June 2, 2021 [1]. Article 109 says it was prepared in line with EU Regulation 2017/745, the EU MDR [1].

Three parts of the regulation matter for a US plan:

  • CE marking. Article 20 requires devices, other than custom-made and investigational devices, to carry the CE marking [1]. Your Turkish market access is EU-style conformity assessment, often through a notified body.
  • EUDAMED and UDI. Articles 27 and 28 point to the European Commission's UDI system and database, not a Turkish one [1].
  • ÜTS. Article 34 requires economic operators, health institutions and others established in Turkey to meet their ÜTS obligations, as set by TİTCK [1].

ÜTS replaced the older TİTUBB database. TİTCK opened ÜTS for medical device registration on June 12, 2017 [3]. Turkey's transition dates for old directive certificates are in our complete 510(k) path for Turkish manufacturers.

Does FDA recognize Turkey, or does Turkey recognize FDA?

No, not for normal market access in either direction. FDA has no route that accepts a Turkish decision, and Turkey does not accept an FDA clearance in place of CE marking.

On the US side, FDA lists mutual recognition agreements with the EU, Switzerland and the UK. They let drug inspectors rely on each other's inspections. FDA's page does not mention medical devices or Turkey [9].

On the Turkish side, the Tıbbi Cihaz Yönetmeliği does not mention FDA. It builds market access on CE marking [1]. So an FDA-cleared device still needs CE marking to be sold in Turkey.

The one link we found is narrow. Under Article 59, TİTCK can let a specific device on the market without the usual conformity assessment when public health or patient safety needs it. TİTCK's guide to these requests asks for the device's FDA status, if it has one, among the supporting documents. But the device must still have a complete MDR technical file. An exemption lasts at most 6 months, with up to two 6-month extensions [1].

Keep your CE marking either way: the two files serve different markets. If you are deciding which file to build first, see FDA first or CE mark first.

What transfers from a Turkish technical file to a 510(k)?

Objective evidence transfers. Certificates and regulatory conclusions do not.

The table goes item by item. "Reusable" means the item can usually go into a 510(k) after a check. "Partly" means some content carries over and some must be redone.

Item in your technical fileAt FDAWhy
Bench, electrical safety and EMC test reportsReusable or partlyFDA looks at the standard and version you tested to, not where the lab is. Check each edition in FDA's Recognized Consensus Standards database.
Biocompatibility (ISO 10993)PartlyFDA applies ISO 10993-1 through its own guidance, last issued September 2023 [8]. In May 2026 FDA recognized the 2025 edition of ISO 10993-1 in part. Endpoints and reporting may need to be added.
Clinical dataPartlyFDA can accept studies run outside the US if they followed good clinical practice and meet 21 CFR 812.28 [5]. A 510(k) usually needs clinical data only when non-clinical testing cannot show substantial equivalence.
Clinical Evaluation ReportNoIt answers an EU question. A 510(k) argues substantial equivalence to a US predicate.
Risk management file (ISO 14971)ReusableThe QMSR incorporates ISO 13485:2016 by reference [6], and that standard builds risk management into product realization. Your risk file is the base for the risk content in the submission.
ISO 13485 certificatePartlyThe quality system aligns with the QMSR. But FDA says a certificate "will not exempt a manufacturer from an FDA inspection" [6]. An MDSAP audit report can replace a routine FDA inspection [7].
Labeling and instructions for usePartlyContent carries over. Required statements must meet 21 CFR Part 801 and appear in English, or as symbols that meet 801.15(c) [5].
UDI dataPartlyThe label must bear a UDI under 21 CFR 801.20, and the labeler submits device data to FDA's GUDID [5, 10]. EUDAMED or ÜTS data does not fill GUDID.
Notified body certificate and CE declaration of conformityNoFDA does not use either one.
ÜTS recordNoIt is a Turkish market record. FDA has its own registration and listing.
EU equivalent deviceNoA 510(k) predicate must be legally marketed in the US.

How is ÜTS different from FDA registration, listing and GUDID?

ÜTS does in one Turkish system what FDA splits into three. FDA registration records the site, listing records the devices, and GUDID records the device identifiers.

ÜTS (Turkey)FDA establishment registrationFDA device listingFDA GUDID
What it isTİTCK's tracking system for medical devices and cosmetics [2]Annual record of each site that makes or processes devices for the US [4]Record of each device a registered site sells in the US [4]FDA's catalog of devices that carry a UDI [10]
Who registersFirms established in Turkey, through Turkey's trade registry (MERSİS) or tax system (VEDOP). A foreign maker's devices are registered by its Turkish importer [1, 3]Owner or operator of each site, foreign or domestic [4]The same owner or operator [4]The labeler [5]
What is recordedCompany, documents such as the EC certificate, declaration of conformity and label, and each product at barcode level [3]Site, activities and, for a foreign site, one US Agent [4]Device by product code and, if needed, its 510(k) or PMA number [4]The device identifier part of the UDI and device data, not production identifiers [10]
TrackingUnit or lot tracking and recalls, as TİTCK decides [2]NoneNoneNone
WhenRequired of device manufacturers and importers in Turkey [3]Within 30 days of starting operations, renewed October 1 to December 31 each year [4]Before import for a foreign site [4]For each device that must carry a UDI [5]

How many Turkish devices has FDA cleared?

FDA cleared 28 510(k)s for Turkey-based applicants from January 1, 2023 to November 26, 2026, by our count of openFDA data [11]. All 28 were Traditional 510(k)s, and they went to 18 companies once spelling variants of the same company name are merged.

General hospital devices led with 8 clearances, then orthopedic with 7 and dental with 4 [11]. The count uses the applicant's address, not the manufacturing site. A Turkish company that files through a US subsidiary is counted under the US. The five most recent clearances were:

K numberDeviceApplicantDecision date
K262091FRAXION Spinal Fixation SystemOrtler Global Tibbi Gerecler Anonim SirketiSeptember 17, 2026
K261830FRAXION plate and bone screw systemsOrtler Global Tibbi Gerecler Anonim SirketiAugust 28, 2026
K260277Mikron trauma bone plate and screw systemMikron Makina San. ve Tic. Ltd. Sti.August 13, 2026
K262681NanoCeram Resin (dental)3BFAB Teknoloji A.S.July 31, 2026
K260234MSK Go (radiology software)Smart Alfa Teknoloji San. Ve Tic. A.S.May 21, 2026

openFDA's registration data shows 92 Turkey-based establishments, 77 of them registered as manufacturers. Of those 77, 50 list no device with a 510(k) or PMA number [11]. So for many Turkish makers, the US route is registration, listing and a US Agent, with no 510(k). Our guide to selling Class I devices in the US from Turkey walks through each step.

What should Turkish manufacturers do first?

Start with the FDA product code, because it decides everything else. Your Turkish or EU risk class does not convert into an FDA class.

  1. Find your product code and class. Search FDA's Product Classification database, or answer six questions in our class and pathway finder.
  2. Check whether the device is 510(k) exempt. In FDA's classification data, 2,155 of 2,403 Class I product codes are 510(k) exempt [11]. Exemptions have limits. Under 21 CFR 880.9, for example, a device still needs a 510(k) if it has a different intended use [5]. The same applies if it uses a different fundamental scientific technology. An exempt device goes straight to registration, listing and a US Agent. Our team handles establishment registration, device listing and the US Agent role in one engagement.
  3. If you need a 510(k), choose a US predicate. It must be legally marketed in the US. A device sold only in Turkey or the EU cannot serve.
  4. Map your test reports. Compare each report with the FDA-recognized version of its standard, and with FDA's ISO 10993-1 guidance for biocompatibility [8]. Plan only the tests you are missing.
  5. Prepare US labeling and UDI. Write labeling to 21 CFR Part 801 in English, and plan your GUDID records [5, 10].
  6. Register and list before the first shipment. A foreign site must list its devices before they are imported or offered for import, and it must name one US Agent [4]. This comes after clearance, not before.

If you want a second view on steps 1 to 3, use our pathway assessment. It gives your device's likely FDA class, candidate product codes and the predicate devices you could file against. Complizen has its own team in Turkey, working alongside our US team.

Frequently asked questions

Does FDA recognize a TİTCK registration?

No. FDA makes its own decision on each device and has no route that relies on a TİTCK decision or an ÜTS record. The evidence behind your Turkish file can still help, such as test reports to FDA-recognized standards and an ISO 13485 quality system, but FDA reviews it again.

Does FDA accept the CE mark in place of a 510(k)?

No. A CE mark has no legal standing at FDA, so a device that needs a 510(k) still needs one. Many reports in a CE technical file can be reused, but the Clinical Evaluation Report, notified body certificate and EU equivalent device do not transfer to a 510(k).

Is there a US and Turkey mutual recognition agreement for medical devices?

We found none. FDA lists mutual recognition agreements with the EU, Switzerland and the UK, and they cover drug inspections, not medical devices. FDA's page, current as of March 28, 2025, does not mention Turkey. Check FDA's page again before relying on this.

Can I sell an FDA-cleared device in Turkey without CE marking?

No. Turkey's Medical Device Regulation requires CE marking, other than for custom-made and investigational devices, and does not mention FDA clearance. TİTCK can grant a short, case-by-case exemption under Article 59 for public health needs, and its guide asks for the device's FDA status as supporting evidence.

Is ÜTS the same as FDA registration?

No. ÜTS is TİTCK's tracking system for devices sold in Turkey, and it records companies, documents and products at barcode level. FDA splits this work into establishment registration, device listing and GUDID. Being in ÜTS does not register you with FDA.

Can I use test reports from a Turkish lab in a 510(k)?

Often, yes. FDA looks at the standard and version you tested to and how complete the report is, not where the lab is. EU harmonized editions can differ from the editions FDA recognizes, so check each one in FDA's Recognized Consensus Standards database before you plan new tests.

Does an ISO 13485 certificate replace an FDA inspection?

No. FDA says a certificate of conformance to ISO 13485 will not exempt a manufacturer from an FDA inspection. Your quality system still aligns with the QMSR, which took effect February 2, 2026. An MDSAP audit report is different, because FDA may accept it in place of a routine inspection.

Can my EU equivalent device be my 510(k) predicate?

Only if it is also legally marketed in the US. A 510(k) predicate must be a US device, such as one already cleared through a 510(k). A device sold only in Turkey or the EU cannot serve as a predicate, even if it is the same product.

Do Class I devices from Turkey need a 510(k)?

Most do not. In FDA's classification data, 2,155 of 2,403 Class I product codes are 510(k) exempt. Exemptions have limits, such as a different intended use, and some Class I devices still need a 510(k). Every device still needs FDA registration, listing and a US Agent before import.

When does a Turkish manufacturer need a US Agent?

Before the device is imported or offered for import into the US, not before 510(k) clearance. Each foreign site names one US Agent, who lives or has a place of business in the US. Your EU Authorised Representative or Turkish importer cannot fill this role unless it meets that rule.

Do I need GUDID if my device is already in ÜTS?

Yes, if the device must carry a UDI under FDA rules. The labeler submits device data to GUDID, FDA's own database, and ÜTS or EUDAMED records do not count. FDA has exceptions to the UDI rules, so check whether one applies to your device.

Should I drop CE marking to focus on FDA?

No. CE marking is how you sell in Turkey and the EU, and FDA clearance is how you sell in the US. Much of the same evidence supports both. The useful question is which file to build first, which depends on your markets and your device.

References

  1. Türkiye İlaç ve Tıbbi Cihaz Kurumu. Tıbbi Cihaz Yönetmeliği (Articles 20, 27, 28, 34, 59, 109). Published in the Resmî Gazete No. 31499 (Mükerrer), June 2, 2021; amended in Resmî Gazete No. 31907 (July 29, 2022), No. 32151 (April 2, 2023) and No. 32635 (August 17, 2024). With TİTCK's guide to Article 59 exemptions (TCOKKA-KLVZ-03), in force August 12, 2023. Accessed November 27, 2026.
  2. Türkiye İlaç ve Tıbbi Cihaz Kurumu. Ürün Takip Sistemi (ÜTS). Accessed November 27, 2026.
  3. Türkiye İlaç ve Tıbbi Cihaz Kurumu. Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz (TCKKD-KLVZ-03), first version in force May 13, 2026 (Articles 4 and 5). With TİTCK's announcement on the opening of ÜTS medical device registration screens, June 2017.
  4. eCFR. 21 CFR Part 807, Establishment Registration and Device Listing (§§ 807.22, 807.28, 807.39, 807.40). Current as of November 2026.
  5. eCFR. 21 CFR 801.15, 801.20, 812.28, 830.300 and 880.9. Current as of November 2026.
  6. FDA. Quality Management System Regulation: Frequently Asked Questions. Content current as of November 5, 2026.
  7. FDA. Medical Device Single Audit Program (MDSAP). Content current as of December 19, 2025.
  8. FDA. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1". Final guidance, September 2023.
  9. FDA. Mutual Recognition Agreements (MRA). Content current as of March 28, 2025.
  10. FDA. Global Unique Device Identification Database (GUDID). Content current as of December 17, 2024.
  11. openFDA. 510(k) API and Registration and Listing API (country code TR), and Device Classification API. Data updated November 16, 2026; pulled November 27, 2026.